TREK™ Coronary Dilatation Catheter’s unique design provides smooth transitions from hub to tip.
MINI TREK™ Coronary Dilatation Catheter’s ultra low profile design enables lesion access.
Skive design provides smooth transition between hypotube and distal shaft.
Confident crossing in resistant lesions.
Provides reduced wall thickness for flexibility and low tip crossing profiles.
Radiopaque flexible markers for confident placement.
Thin, multi-layer balloon technology provides low crossing profiles for access to complex lesions and a tri-fold design†† for excellent re-wrap.
TREK™ & MINI TREK™ Coronary Dilatation Catheters | |
---|---|
Nominal Pressure (NP) | 8 atm |
Rated Burst Pressure (RBP) | 14 atm |
Tip Entry Profile | 0.017 in (3.00 mm Ø) |
Tip Crossing Profile | 0.021 in average (3.00 mm Ø) 0.023 in maximum |
Folded Balloon Crossing Profile | 0.032 in average (3.00 mm Ø) 0.036 in maximum (3.00 mm Ø) |
Refolded Balloon Crossing Profile | 0.056 in maximum (3.00 mm Ø) |
RX Notch Diameter (For RX Platform) | 2.5 French average (3.00 mm Ø) 2.7 French maximum (all Ø) |
Tip Length | <3 mm |
Balloon Material | Pebax†, semi-compliant |
Marker Material | Flexible Tungsten/Pebax‡ |
Catheter Length | 145 cm |
Balloon Coating | Hydrophilic |
Minimum Guide Catheter | 5 French (1.20 mm – 4.00 mmØ) 6 French (4.50 mm & 5.00 mmØ) |
Maximum Guide Wire Compatibility | 0.014 in |
Latex Free | Yes |
Balloon Diameter (mm) | Balloon Length (mm) | |||||||
---|---|---|---|---|---|---|---|---|
6 | 8 | 12 | 15 | 20 | 25 | 30 | ||
MINI TREK™ RX (CTO INDICATED) | 1.20 | 1012268-06U | 1012268-08U | 1012268-12U | 1012268-15U | 1012268-20U | - | - |
1.50 | 1012269-06 | 1012269-08 | 1012269-12 | 1012269-15 | 1012269-20 | - | - | |
2.00 | 1012270-06 | 1012270-08 | 1012270-12 | 1012270-15 | 1012270-20 | 1012270-25 | 1012270-30 | |
TREK RX™ | 2.25 | 1012271-06 | 1012271-08 | 1012271-12 | 1012271-15 | 1012271-20 | 1012271-25 | 1012271-30 |
2.50 | 1012272-06 | 1012272-08 | 1012272-12 | 1012272-15 | 1012272-20 | 1012272-25 | 1012272-30 | |
2.75 | 1012273-06 | 1012273-08 | 1012273-12 | 1012273-15 | 1012273-20 | 1012273-25 | 1012273-30 | |
3.00 | 1012274-06 | 1012274-08 | 1012274-12 | 1012274-15 | 1012274-20 | 1012274-25 | 1012274-30 | |
3.25 | 1012275-06 | 1012275-08 | 1012275-12 | 1012275-15 | 1012275-20 | 1012275-25 | 1012275-30 | |
3.50 | 1012276-06 | 1012276-08 | 1012276-12 | 1012276-15 | 1012276-20 | 1012276-25 | 1012276-30 | |
3.75 | 1012277-06 | 1012277-08 | 1012277-12 | 1012277-15 | 1012277-20 | 1012277-25 | 1012277-30 | |
4.00 | 1012278-06 | 1012278-08 | 1012278-12 | 1012278-15 | 1012278-20 | 1012278-25 | 1012278-30 | |
4.50 | - | - | 1012279-12 | 1012279-15 | - | - | - | |
5.00 | - | - | 1012280-12 | 1012280-15 | - | - | - |
Balloon Diameter (mm) | Balloon Length (mm) | |||||||
---|---|---|---|---|---|---|---|---|
6 | 8 | 12 | 15 | 20 | 25 | 30 | ||
MINI TREK™ II OTW (CTO INDICATED) | 1.20 | 1012401-06Z | 1012401-08Z | 1012401-12Z | 1012401-15Z | 1012401-20Z | - | - |
1.50 | 1012402-06A | 1012402-08A | 1012402-12A | 1012402-15A | 1012402-20A | - | - | |
2.00 | 1012403-06A | 1012403-08A | 1012403-12A | 1012403-15A | 1012403-20A | - | - | |
TREK OTW™ | 2.25 | 1012404-06 | 1012404-08 | 1012404-12 | 1012404-15 | 1012404-20 | 1012404-25 | 1012402-30 |
2.50 | 1012405-06 | 1012405-08 | 1012405-12 | 1012405-15 | 1012405-20 | 1012405-25 | 1012405-30 | |
2.75 | 1012406-06 | 1012406-08 | 1012406-12 | 1012406-15 | 1012406-20 | 1012406-25 | 1012406-30 | |
3.00 | 1012407-06 | 1012407-08 | 1012407-12 | 1012407-15 | 1012407-20 | 1012407-25 | 1012407-30 | |
3.25 | 1012408-06 | 1012408-08 | 1012408-12 | 1012408-15 | 1012408-20 | 1012408-25 | 1012408-30 | |
3.50 | 1012409-06 | 1012409-08 | 1012409-12 | 1012409-15 | 1012409-20 | 1012409-25 | 1012409-30 | |
3.75 | 1012410-06 | 1012410-08 | 1012410-12 | 1012410-15 | 1012410-20 | 1012410-25 | 1012410-30 | |
4.00 | 1012411-06 | 1012411-08 | 1012411-12 | 1012411-15 | 1012411-20 | 1012411-25 | 1012411-30 | |
4.50 | - | - | 1012412-12 | 1012412-15 | - | - | - | |
5.00 | - | - | 1012413-12 | 1012413-15 | - | - | - |
Inflation Pressure (atm) | Balloon Diameter (mm) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1.20 | 1.50 | 2.00 | 2.25 | 2.50 | 2.75 | 3.00 | 3.25 | 3.50 | 3.75 | 4.00 | 4.50 | 5.00 | |
MINI TREK™ | TREK™ | ||||||||||||
2 | 1.10 | 1.34 | 1.69 | 1.87 | 2.15 | 2.16 | 2.49 | 2.91 | 2.93 | 3.06 | 3.24 | 3.69 | 4.10 |
3 | 1.12 | 1.36 | 1.80 | 1.95 | 2.20 | 2.35 | 2.56 | 2.98 | 3.04 | 3.22 | 3.46 | 3.91 | 4.28 |
4 | 1.13 | 1.38 | 1.85 | 2.00 | 2.27 | 2.41 | 2.66 | 3.05 | 3.11 | 3.31 | 3.56 | 4.04 | 4.43 |
5 | 1.14 | 1.40 | 1.90 | 2.06 | 2.34 | 2.49 | 2.73 | 3.12 | 3.20 | 3.41 | 3.67 | 4.17 | 4.59 |
6 | 1.17 | 1.43 | 1.94 | 2.12 | 2.40 | 2.56 | 2.81 | 3.19 | 3.28 | 3.50 | 3.78 | 4.29 | 4.73 |
7 | 1.18 | 1.45 | 1.98 | 2.17 | 2.45 | 2.63 | 2.87 | 3.25 | 3.38 | 3.59 | 3.89 | 4.40 | 4.86 |
8 NP | 1.20 | 1.47 | 2.01 | 2.21 | 2.51 | 2.70 | 2.93 | 3.31 | 3.45 | 3.67 | 3.99 | 4.49 | 4.98 |
9 | 1.21 | 1.49 | 2.03 | 2.25 | 2.55 | 2.75 | 2.97 | 3.36 | 3.51 | 3.74 | 4.06 | 4.56 | 5.05 |
10 | 1.23 | 1.51 | 2.07 | 2.28 | 2.58 | 2.80 | 3.01 | 3.39 | 3.56 | 3.79 | 4.11 | 4.62 | 5.13 |
11 | 1.24 | 1.52 | 2.09 | 2.31 | 2.60 | 2.83 | 3.04 | 3.42 | 3.60 | 3.84 | 4.18 | 4.68 | 5.20 |
12 | 1.25 | 1.54 | 2.12 | 2.33 | 2.62 | 2.86 | 3.07 | 3.45 | 3.64 | 3.87 | 4.23 | 4.73 | 5.25 |
13 | 1.26 | 1.55 | 2.14 | 2.35 | 2.64 | 2.88 | 3.09 | 3.48 | 3.67 | 3.91 | 4.28 | 4.80 | 5.31 |
14 RBP | 1.27 | 1.56 | 2.17 | 2.37 | 2.67 | 2.91 | 3.11 | 3.50 | 3.70 | 3.94 | 4.32 | 4.86 | 5.36 |
15 | 1.28 | 1.57 | 2.19 | 2.39 | 2.69 | 2.92 | 3.14 | 3.53 | 3.73 | 3.99 | 4.38 | 4.92 | 5.41 |
16 | 1.29 | 1.59 | 2.21 | 2.41 | 2.71 | 2.96 | 3.16 | 3.56 | 3.75 | 4.02 | 4.44 | 4.98 | 5.50 |
17 | 1.30 | 1.60 | 2.24 | 2.43 | 2.73 | 2.99 | 3.18 | 3.58 | 3.78 | 4.05 | 4.49 | 5.04 | 5.56 |
Data on file at Abbott unless otherwise noted.
**Slim Seal technology available on 2.0 mm – 3.75 mm balloon sizes.
† Single marker on all 6 mm lengths and 1.20 – 1.50 mm diameters (all lengths). Dual markers on all other sizes.
*CrossFlex2 technology available on 2.25 mm – 5.0 mm balloon sizes.
††Tri-fold design for 2.00 mm – 4.00 mm diameters. Bi-fold design for 1.20 mm & 1.50 mm diameters.
CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE. OBSERVE ALL WARNINGS AND PRECAUTIONS NOTED THROUGHOUT THESE INSTRUCTIONS. FAILURE TO DO SO MAY RESULT IN COMPLICATIONS.
Applies to TREK™ RX & OTW 2.25 mm – 5.00 mm sizes only:
The TREK™ RX & OTW Coronary Dilatation Catheters are indicated for:
Applies to MINI TREK™ RX and MINI TREK™ II OTW 1.50 mm – 2.00 mm sizes only:
The TREK™ RX & OTW Coronary Dilatation Catheters are indicated for:
Applies to MINI TREK™ RX and MINI TREK™ II OTW 1.20 mm sizes only:
The MINI TREK™ RX & MINI TREK™ II OTW 1.20mm Coronary Dilatation Catheters are indicated for:
Note: (applies to 2.00 mm to 5.00 mm only): Post-deployment stent expansion testing was performed on the bench with the MULTILINK VISION™ and MULTI-LINK ULTRA™ stents. All stents should be deployed in accordance with the manufacturer’s indications and instructions for use.
The TREK™ RX & OTW, MINI TREK™ RX and MINI TREK™ II OTW Coronary Dilatation Catheters are not intended to be used to treat patients with:
This device is intended for one time use only. DO NOT resterilize and / or reuse it, as this can compromise device performance and increase the risk of cross contamination due to inappropriate reprocessing.
Percutaneous transluminal coronary angioplasty (PTCA) should only be performed at hospitals where emergency coronary artery bypass graft surgery can be quickly performed in the event of a potentially injurious or lifethreatening complication.
PTCA in patients who are not acceptable candidates for coronary artery bypass graft surgery requires careful consideration, including possible hemodynamic support during PTCA, as treatment of this patient population carries special risk.
Use only the recommended balloon inflation medium. Never use air or any gaseous medium to inflate the balloon.
Balloon pressure should not exceed the rated burst pressure (RBP). The RBP is based on results of in vitro testing. At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their RBP. Use of a pressuremonitoring device is recommended to prevent over pressurization.
To reduce the potential for vessel damage, the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the stenosis.
When the catheter is exposed to the vascular system, it should be manipulated while under high quality fluoroscopic observation.
Do not advance or retract the catheter unless the balloon is fully deflated under vacuum. If resistance is met during manipulation, determine the cause of the resistance before proceeding.
Do not use, or attempt to straighten, a catheter if the shaft has become bent or kinked; this may result in the shaft breaking. Instead, prepare a new catheter.
Do not torque the catheter more than one (1) full turn.
Treatment of moderately or heavily calcified lesions is considered to be moderate risk, with an expected success rate of 60 – 85% and increases the risk of acute closure, vessel trauma, balloon burst, balloon entrapment, and associated complications. If resistance is felt, determine the cause before proceeding. Continuing to advance or retract the catheter while under resistance may result in damage to the vessels and / or damage / separation of the catheter.
In the event of catheter damage / separation, recovery of any portion should be performed based on physician determination of individual patient condition and appropriate retrieval protocol.
Note the “Use by” date specified on the package.
Inspect all product prior to use. Do not use if the package is open or damaged.
This device should be used only by physicians trained in angiography and PTCA, and / or percutaneous transluminal angioplasty (PTA).
Prior to angioplasty, the dilatation catheter should be examined to verify functionality and ensure that its size is suitable for the specific procedure for which it is to be used.
During the procedure, appropriate anticoagulant and coronary vasodilator therapy must be provided to the patient as needed. Anticoagulant therapy should be continued for a period of time to be determined by the physician after the procedure.
If the surface of the TREK™ RX & OTW, MINI TREK™ RX or MINI TREK™ II OTW Coronary Dilatation Catheter becomes dry, wetting with heparinized normal saline will reactivate the coating.
Do not reinsert the TREK™ RX & OTW, MINI TREK™ RX or MINI TREK™ II OTW Coronary Dilatation Catheter into the coil dispenser after procedural use.
The safety and effectiveness of this PTCA balloon catheter for the treatment of in-stent restenosis (ISR) have not been established.
Applies to TREK™ RX and MINI TREK™ RX only (APPLIES TO ALL SIZES), in addition to above:
The design and construction of these catheters do not provide the user with distal pressure monitoring capability.
Applies to TREK™ RX 4.50mm and 5.00mm sizes only, in addition to above:
With 4.5 mm and 5.0 mm balloon dilatation catheters, some increased resistance may be noted upon insertion or withdrawal into or out of the guiding catheter. Choosing a larger guiding catheter size may minimize this.
Applies to TREK™ OTW and MINI TREK™ II OTW (APPLIES TO ALL SIZES), in addition to above:
Bench testing was conducted with 0.014” (.36mm) constant diameter guide wires to establish guide wire compatibility. If another type of guide wire is selected with a different dimensional profile, the compatibility (e.g., wire resistance) should be considered prior to use.
Possible adverse effects include, but are not limited to, the following:
MAT-2200916 v1.0