Our manuals contain the latest instructions and information for cardiac physicians, radiologists and MR technologists. We recommend that you download and review the most current version any time a patient with an Abbott product needs an MRI.
Use this database to search for any cardiovascular instructions for use. These include arrhythmia, heart failure, structural heart, biliary and mitral valve repair and other vascular IFUs. You can search by product name, model number, category or family.
Use this database for coronary and peripheral orbital atherectomy system IFUs and their related products.
To look up your patient’s system, use the Product Search tool to determine whether your patient’s device and leads (if applicable) are safe for an MRI.
Based on the country you select, you can choose the device and lead combination (if applicable) to get scan parameters. If a device is not shown in the list, it is not MR Conditional. Alternatively, you can also use the latest MRI Manuals to determine scan parameters.
Download the MRI Ready Systems Manual or MRI Ready Leadless System Manual to review a list of adverse conditions and effects.
CAUTION: Do not bring any external control devices, such as the Merlin™ Patient Care System (PCS) Programmer, SJM MRI Activator™ handheld device, or AVEIR™ Link Module into the scanner magnet room (Zone IV). These devices are considered MR Unsafe.
The MRI parameter settings are selected at the physician's discretion.
The default MRI parameter settings are automatically stored in the Abbott MR Conditional device. Download the MRI Ready Systems Manual or MRI Ready Leadless System Manual for details on default settings and how to save modified settings for scanning.
Follow the checklist instructions within Merlin Programmer. Download the MRI Ready Systems Manual or MRI Ready Leadless System Manual to see examples of checklist screens and get additional details about the checklist.
Once you have completed the checklist, select the Program MRI Settings button to enable MRI Settings.
It is required to program the device to MRI Settings as part of the MRI scan workflow.
Immediately following the MRI procedure, follow the instructions in the MRI Ready Systems Manual or MRI Ready Leadless System Manual to revert to the permanently programmed settings.
To look up your patient’s system, use the Product Search tool to determine whether your patient’s device and leads (if applicable) are safe for an MRI.
Based on the country you select, you can choose the device and lead combination (if applicable) to get scan parameters. If a device is not shown in the list, it is not MR Conditional. Alternatively, you can also use the latest MRI Manuals to determine scan parameters.
Don’t scan the patient if any adverse conditions are present. Download the MRI Ready Systems Manual or MRI Ready Leadless System Manual to review a list of adverse conditions and effects.
Download the MRI Ready Systems Manual or MRI Ready Leadless System Manual to check the MRI scan parameter tables for the applicable scan and identify the settings for RF Transmit Conditions and scan regions.
Be sure to check the lead combination(s) (if applicable) to ensure you select the correct settings.
CAUTION: Do not bring any external control devices, such as the Merlin Patient Care System (PCS) Programmer, SJM MRI Activator handheld device, or AVEIR Link Module into the scanner magnet room (Zone IV). These devices are considered MR Unsafe.
Download the MRI Ready Systems Manual or MRI Ready Leadless System Manual for details about checking the MRI Settings status. For certain pacemakers, the MRI Activator handheld device may be used.
Proper patient monitoring must be provided during the MRI scan. This includes continuous monitoring of the patient's hemodynamic function.
Download the MRI Ready Systems Manual or MRI Ready Leadless System Manual for additional details about patient monitoring.
MRI Settings must be disabled by the patient's device management physician or clinician using the Merlin Programmer, and the AVEIR Link Module for leadless pacemakers. For certain pacemakers, the MRI Activator handheld device may be used.