The AmplatzerTM Family of Vascular Plugs


 

Advancing the Standard of Care in Peripheral Embolisation

  • A single device solution that allows for faster procedure times, less radiation exposure and lower procedural costs1, 2
  • Rapid embolisation even in high flow vessels2-4
  • Designed for precise placement, controlled delivery and fully recapturable
  • A family of plugs for different vessel conditions
  • Limited imaging artifact5

AmplatzerTM Vascular Plugs (AVP) Mechanism of Action

The nitinol mesh of an AVP acts as an embolic agent that promotes clot formation. Each AVP creates a physical barrier to blood flow by instigating both the body’s immune response followed by a natural coagulation process.6

Intended Purpose7

The AmplatzerTM Vascular Plug 4 is intended for therapeutic embolisation to reduce or obstruct blood flow.

Indication for Use8

The AmplatzerTM Family of Vascular Plugs are indicated for arterial and venous embolisation in the peripheral vasculature.

AmplatzerTM Vascular Plug (AVP)

Designed to provide precise positioning in short landing zones6

  • Single lobe, single layer device8
  • Platinum marker bands on both ends8
  • Select a device that is ~30-50% larger in diameter than the target vessel diameter8

Device Description:8

  • Self-expanding device made from a Nitinol wire mesh
  • The device is shipped attached to a 135 cm nitinol delivery wire in a hoop dispenser
  • The device is preloaded in a loader
  • No plastic vice provided
  • Size ranging from 4 – 16 mm in 2 mm increments

Materials:8

 

  • Plug: Nitinol
  • Delivery wire: Nitinol
  • Marker bands: Platinum
  • Micro screw: Stainless steel

Sizing and Device Selection:

A: Vascular plug diameter

B: Unconstrained length

Vessel SizeDevice DimensionsDelivery Catheter Requirements*
Treatable Vessel Diameter RangeModel/Reorder NumberVascular Plug Diameter (A)Unconstrained Length (B)Minimum and Maximum Internal DiameterMinimum Sheath SizeorMinimum Guide Catheter SizeMaximum Delivery Catheter Length**
2.5 mm-
3.0 mm
9-PLUG-0044 mm7 mm≥1.42 mm / ≥0.056 in≥4 F ≥5 F≤ 100 cm
4.0 mm-
4.5 mm
9-PLUG-0066 mm7 mm≥1.42 mm / ≥0.056 in≥4 F ≥5 F≤ 100 cm
5.5 mm-
6.0 mm
9-PLUG-0088 mm7 mm≥1.42 mm / ≥0.056 in≥4 F ≥5 F≤ 100 cm
6.5 mm-
7.5 mm
9-PLUG-01010 mm7 mm≥1.68 mm / ≥0.066 in≥5 F ≥6 F≤ 100 cm
8.0 mm-
9.0 mm
9-PLUG-01212 mm8 mm≥1.68 mm / ≥0.066 in≥5 F ≥6 F≤ 100 cm
9.5 mm-
11.0 mm
9-PLUG-01414 mm8 mm≥2.21 mm / ≥0.087 in≥6 F ≥8 F≤ 100 cm
10.5 mm-
12.5 mm
9-PLUG-01616 mm8 mm≥2.21 mm / ≥0.087 in≥6 F ≥8 F≤ 100 cm


 

Treatable vessel diameter range based on the devices Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
*The Amplatzer™ Vascular Plug is delivered utilising either a sheath or guide catheter meeting the minimum internal diameter requirements above.
**Each Amplatzer™ Vascular Plug comes pre-loaded on a 135 cm nitinol delivery wire.

Deployment

Note: This section should not be used in place of the Instructions for Use. This section only provides an overview of the procedure. Refer to the Instructions for Use for additional information.8

Access the vessel and perform an angiogram using standard technique to measure the diameter of the vessel at the desired occlusion site.

Select a vascular plug with a diameter approximately 30-50% larger than the vessel diameter at the occlusion site. Ensure that the occlusion site has sufficient length so the deployed device length will not obstruct other vessels or anatomical structures.

Deployment steps

1Flush hoop dispenser and loader with sterile saline until fluid exits the distal tip to purge air from the loader.
2Remove device (in loader) and delivery wire from the hoop dispenser.
3Select a delivery catheter with an internal diameter that will accommodate the selected device. The original access catheter may be used if the internal diameter is sufficient for the selected device. Prepare the catheter for use (according to the manufacturer’s instructions for use).
4Advance the delivery catheter over the guidewire until the distal tip of the catheter is at the distal edge of the occlusion site (according to the manufacturer’s instructions for use).
5Remove the guidewire.
6Insert loader into the delivery catheter through a Y-connector or hemostasis valve.
7Allow blood backflow or aspirate the system to ensure air is purged from the catheter and loader.
8Do not over tighten Y-connector to avoid damaging loader.
9Advance the delivery wire to advance the device to the distal end of the delivery catheter. Do not twist or rotate the delivery wire during advancement to ensure device does not prematurely detach.
10Hold delivery wire in place and slowly retract the delivery catheter to deploy the device at the occlusion site.
11Verify position of the device using the radiopaque marker bands.
~ If unsatisfactory, stabilise the wire and re-advance the delivery catheter until the device is completely within the catheter. Reposition and deploy, or remove the device from the patient.
~ If satisfactory, attach the plastic visea to the wire and rotate the delivery wire counterclockwise until it separates from the device.
12Remove delivery catheter and wire from the patient.
Steps 1 - 3
Steps 4-5
Steps 6-9
Steps 10-12

AmplatzerTM Vascular Plug II (AVP II)

Designed for rapid embolisation2 and can adjust to variable landing zones9

  • Multi-layered mesh designed to increase density and flow disturbance10
  • Six planes of cross-sectional coverage8 resulting in timely occlusion2
  • Select a device that is ~30 - 50% larger in diameter than the target vessel diameter10

Device Description:10

  • Self-expanding device made from a Nitinol wire mesh
  • The device is shipped attached to a 135 cm nitinol delivery wire in a hoop dispenser
  • The device is preloaded in a loader
  • A plastic vise is also included and may be attached to the delivery wire to facilitate device detachment

Materials:10

 

  • Plug: Nitinol
  • Delivery wire: Nitinol
  • Marker bands: Platinum
  • Micro screw: Stainless steel

Sizing and Device Selection:

 

A: Vascular plug diameter

B: Unconstrained length

Vessel SizeDevice DimensionsDelivery Catheter Requirements*
Treatable Vessel Diameter RangeModel/ Reorder NumberVascular Plug Diameter (A)Unconstrained Length (B)Minimum and Maximum Internal DiameterMinimum Sheath SizeorMinimum Guide Catheter SizeMaximum Delivery Catheter Length**
2.0 mm-2.5 mm9-AVP2-0033 mm6 mm1.42-1.70 mm / 0.056-0.067 in≥4 F ≥5 F≤ 100 cm
2.5 mm-3.0 mm9-AVP2-0044 mm6 mm1.42-2.50 mm / 0.056-0.098 in≥4 F ≥5 F≤ 100 cm
4.0 mm-4.5 mm9-AVP2-0066 mm6 mm1.42-2.50 mm / 0.056-0.098 in≥4 F ≥5 F≤ 100 cm
5.5 mm-6.0 mm9-AVP2-0088 mm7 mm1.42-2.69 mm / 0.056-0.106 in≥4 F ≥5 F≤ 100 cm
6.5 mm-7.5 mm9-AVP2-01010 mm7 mm1.78-2.69 mm / 0.070-0.106 in≥5 F ≥6 F≤ 100 cm
8.0 mm-9.0 mm9-AVP2-01212 mm9 mm1.78-2.69 mm / 0.070-0.106 in≥5 F ≥6 F≤ 100 cm
9.5 mm-11.0 mm9-AVP2-01414 mm10 mm2.18-2.69 mm / 0.086-0.106 in≥6 F ≥8 F≤ 100 cm
10.5 mm-12.5 mm9-AVP2-01616 mm12 mm2.18-2.69 mm / 0.086-0.106 in≥6 F ≥8 F≤ 100 cm
12.0 mm-14.0 mm9-AVP2-01818 mm14 mm2.49-2.69 mm / 0.098-0.106 in≥7 F ≥9 F≤ 100 cm
13.5 mm-15.5 mm9-AVP2-02020 mm16 mm2.49-2.69 mm / 0.098-0.106 in≥7 F ≥9 F≤ 100 cm
14.5 mm-17.0 mm9-AVP2-02222 mm18 mm2.49-2.69 mm / 0.098-0.106 in≥7 F ≥9 F≤ 100 cm


 

Treatable vessel diameter range based on the devices Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
*The AmplatzerTM Vascular Plug II is delivered utilising either a sheath or guide catheter meeting the minimum internal diameter requirements above.
**Each AmplatzerTM Vascular Plug II comes pre-loaded on a 135 cm nitinol delivery wire.

Deployment

Note: This section should not be used in place of the Instructions for Use. This section only provides an overview of the procedure. Refer to the Instructions for Use for additional information.10

Access the vessel and perform an angiogram using standard technique to measure the diameter of the vessel at the desired occlusion site.

Select a vascular plug with a diameter approximately 30-50% larger than the vessel diameter at the occlusion site. Ensure that the occlusion site has sufficient length so the deployed device length will not obstruct other vessels or anatomical structures.

Deployment steps

1Flush hoop dispenser and loader with sterile saline until fluid exits the distal tip to purge air from the loader.
2Remove device (in loader) and delivery wire from the hoop dispenser.
3Select a delivery catheter with an internal diameter that will accommodate the selected device. The original access catheter may be used if the internal diameter is sufficient for the selected device. Prepare the catheter for use (according to the manufacturer’s instructions for use).
4Advance the delivery catheter over the guidewire until the distal tip of the catheter is at the distal edge of the occlusion site (according to the manufacturer’s instructions for use).
5Remove the guidewire.
6Insert loader into the delivery catheter through a Y-connector or hemostasis valve.
7Allow blood backflow or aspirate the system to ensure air is purged from the catheter and loader.
8Do not over tighten Y-connector to avoid damaging loader.
9Advance the delivery wire to advance the device to the distal end of the delivery catheter. Do not twist or rotate the delivery wire during advancement to ensure device does not prematurely detach.
10Hold delivery wire in place and slowly retract the delivery catheter to deploy the device at the occlusion site.
11Verify position of the device using the radiopaque marker bands.
~ If unsatisfactory, stabilise the wire and re-advance the delivery catheter until the device is completely within the catheter. Reposition and deploy, or remove the device from the patient.
~ If satisfactory, attach the plastic visea to the wire and rotate the delivery wire counterclockwise until it separates from the device.
12Remove delivery catheter and wire from the patient.
Steps 1 - 3
Steps 4-5
Steps 6-9
Steps 10-12

Amplatzer™ Vascular Plug 4 (AVP 4)

Low-profile embolisation compatible with 0.038” guide wire-compatible diagnostic catheters7 that are designed to reach distal vasculature through tortuous anatomy.6

  • Simple delivery through an 0.038” guide wire compatible diagnostic catheter7
  • Easy tracking through tortuous anatomy7
  • Extended reach to distal vasculature7
  • Select a device that is ~30-50% larger in diameter than the target vessel diameter7

Device Description:7

  • The device is shipped attached to a 155 cm delivery wire in a hoop dispenser
  • The device is preloaded in a loader
  • 155 cm long PTFE-coated delivery wire
  • A plastic vise is also included and may be attached to the delivery wire to facilitate device detachment.
  • Compatibility with 0.038” diagnostic catheters

Materials:7

  • Plug: Nitinol
  • Delivery wire: PTFE-coated stainless wire coil with NiTi core
  • Marker bands: Platinum
  • Micro screw: Stainless steel

Sizing and Device Selection:

A: Vascular plug diameter

B: Unconstrained length

Vessel SizeDevice DimensionsDiagnostic Catheter Requirements*
Treatable Vessel Diameter RangeModel/Reorder NumberVascular Plug Diameter (A)Unconstrained Length (B)Requirements for Diagnostic Catheters**Maximum Diagnostic Catheter Length***
2.5 mm-3.0 mm9-AVP038-0044 mm10.0 mm0.038" Guidewire-Compatible Diagnostic Catheter≤ 125 cm
3.5 mm-4.0 mm9-AVP038-0055 mm10.5 mm0.038" Guidewire-Compatible Diagnostic Catheter≤ 125 cm
4.0 mm-4.5 mm9-AVP038-0066 mm11.0 mm0.038" Guidewire-Compatible Diagnostic Catheter≤ 125 cm
4.5 mm-5.5 mm9-AVP038-0077 mm12.5 mm0.038" Guidewire-Compatible Diagnostic Catheter≤ 125 cm
5.5 mm-6.0 mm9-AVP038-0088 mm13.5 mm0.038" Guidewire-Compatible Diagnostic Catheter≤ 125 cm


 

Treatable vessel diameter range based on the devices Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
*The AmplatzerTM Vascular Plug 4 is delivered utilising a Diagnostic Catheter meeting the minimum internal diameter requirements above.
**The AmplatzerTM Vascular Plug 4 has been tested for compatibility with the following diagnostic catheters and corresponding lengths:
    5F Dx Catheters: Boston Scientific IMAGER‡ II (100 cm or less) and Merit Medical Impress‡ (125 cm or less)
    4F Dx Catheters: Cordis‡ TEMPO‡ or TEMPO‡ AQUA‡ (100 cm or less)
***Each AmplatzerTM Vascular Plug 4 comes pre-loaded and packaged on an 0.038" x 155cm-long PTFE-coated delivery wire.

Deployment

Note: This section should not be used in place of the Instructions for Use. This section only provides an overview of the procedure. Refer to the Instructions for Use for additional information.7

Access the vessel and perform an angiogram using standard technique to measure the diameter of the vessel at the desired occlusion site.

Select a vascular plug with a diameter approximately 30-50% larger than the vessel diameter at the occlusion site. Ensure that the occlusion site has sufficient length so the deployed device length will not obstruct other vessels or anatomical structures.

Deployment steps

1a) Flush with sterile saline through the stop cock until fluid exits the distal tip to purge air from the loader.
NOTE: If the AVP 4 is inadvertently deployed before use, follow these additional steps prior to use:
b) grasp the spring-loaded tapered loader tubing and the rotating luer in one hand and retract the delivery wire with the other hand.
c) Continue to retract the delivery wire until both the single white marker and the double white markers on the delivery wire are visible. Make sure the device is fully retracted into the loader.
d) Reflush the device if necessary.
2Remove device (in loader) and delivery wire from the hoop dispenser.
3Select a delivery catheter with an internal diameter that will accommodate the selected device. The original access catheter may be used if the internal diameter is sufficient for the selected device. Prepare the catheter for use (according to the manufacturer’s instructions for use).
4Advance the delivery catheter over the guidewire until the distal tip of the catheter is at the distal edge of the occlusion site (according to the manufacturer’s instructions for use).
5Remove the guidewire.
6Insert the tapered tip of the loader into the hub of the 0.038 inch guidewire compatible diagnostic catheter. NOTE: Do not connect the luer through a stopcock, hemostasis valve, Y-connector or other valve. This will result in damage to the loader.
7Grasp the rotating luer and press the luer to the hub of the catheter and rotate clockwise to ensure full engagement of the loader to the catheter hub.
8Advance the delivery wire until the double white marker is as the hemostasis valve.
9Hold delivery wire in place and slowly retract the delivery catheter to deploy the device at the occlusion site.
10Verify position of the device using the radiopaque marker bands. ~ If unsatisfactory, stabilise the wire and re-advance the delivery catheter until the device is completely within the catheter. Reposition and deploy, or remove the device from the patient. ~ If satisfactory, attach the plastic vise to the wire and rotate the delivery wire counterclockwise until it separates from the device.
11Remove delivery catheter and wire from the patient.
Steps 1 - 2
Steps 3-5
Steps 6-8
Steps 9-11


 

Discover Abbott’s Peripheral intervention technology


 

ABBOTT’S PERIPHERAL INTERVENTION PORTFOLIO

References:

1. Hart JL, et al. Eur Radiol; 2010. 20(11): 2663-70.
2. Pech M, et al. Cardiovasc Intervent Radiol; 2009. 32(3): 455-61.
3. Abbott, Testing data on file.
4. Kucukay F, et al. J Vasc Interv Radiol; 2014. 25(9): 1327-32.
5. Vandy F, et al. J Vasc Surg; 2008. 48(5): 1121-4.
6. Lopera JE, Semin Intervent Radiol; 2015. 32(4): 356-69.
7. Abbott, AmplatzerTM Vascular Plug 4 - Instructions for Use.
8. Abbott, AmplatzerTM Vascular Plug - Instructions for use.
9. AVP II is available in 8 unconstrained lengths ranging from 6 -18 mm.
10. Abbott AmplatzerTM Vascular Plug II - Instructions for Use.

 

 

MAT-2117128 V1.0