Nonclinical testing has shown that the XIENCE Skypoint™ and XIENCE Sierra™ Stents, in single and in overlapped configurations up to 91 mm and 71 mm in length respectively, are MR conditional. A patient with a XIENCE™ Stent can be scanned in an MR system under the following conditions:*
Under the scan conditions defined above, XIENCE™ Stents are expected to produce a maximum temperature rise of of less than 4.8°C after 15 minutes of continuous scanning.
In non-clinical testing, the image artifact caused by the device extends approximately 6 mm from the XIENCE™ Stent when imaged with a gradient echo or spin echo pulse sequence and a 3T MRI system. It may be necessary to optimize the MRI parameters to account for the presence of XIENCE™ Stents.
Refer to the Instructions For Use (IFU) for additional information.
XIENCE Skypoint™ Stent offers these features:
The ultra-low crossing profile allows users to choose XIENCE™ Stent for the most challenging cases of coronary artery disease. XIENCE™ Stents, in addition to being the most studied stents, produce consistently low stent thrombosis (ST) rates.1
With more than 30,000 XIENCE™ DES patients analysed in DAPT studies,4 Abbott has received a 1-month DAPT CE indication for High Bleeding Risk patients.2
In addition, the STOPDAPT, STOPDAPT-2, STOPDAPT-3, XIENCE 28 and XIENCE 90 trials reported consistently low ST rates when DAPT was discontinued after 1 or 3 months.4 (XIENCE 28, STOPDAPT, STOPDAPT-2 and XIENCE 90 results at 12 months). Learn more about DAPT.
* The effects of MRI on overlapped stents greater than 71 mm in length, or involving stents with fractured struts, are unknown.
MAT-2413713 V1.0
Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions