Current evidence suggests that percutaneous occlusion of the Left Atrial Appendage (LAA) reduces the risk of thromboembolic complications associated with non-valvular atrial fibrillation.1 The Amplatzer™ Amulet™ LAA Occluder is designed for complete occlusion of the LAA, thereby reducing the risk for ischemic stroke caused by atrial fibrillation.
Since LAA anatomies come in all shapes and sizes, the Amplatzer Amulet LAA Occluder features the broadest size range of occluders on the market today.2
Proximal device positioning allows for proper placement regardless of:
Reorder Number | Amulet Device Size/Lobe Diameter | Landing Zone Width (mm) | Lobe Length (mm) | Minimum Sheath Size (mm) | Minimum LAA Depth (mm) | Disc Diameter (mm) |
---|---|---|---|---|---|---|
9-ACP2-007-016 | 16 | 11.0-13.0 | 7.5 | 12 F | ≥ 10 | 22 |
9-ACP2-007-018 | 18 | 13.0-15.0 | 7.5 | 12 F | ≥ 10 | 24 |
9-ACP2-007-020 | 20 | 15.0-17.0 | 7.5 | 12 F | ≥ 10 | 26 |
9-ACP2-007-022 | 22 | 17.0-19.0 | 7.5 | 12 F | ≥ 10 | 28 |
9-ACP2-010-025 | 25 | 19.0-22.0 | 10 | 12 F | ≥ 12 | 32 |
9-ACP2-010-028 | 28 | 22.0-25.0 | 10 | 14 F | ≥ 12 | 35 |
9-ACP2-010-031 | 31 | 25.0-28.0 | 10 | 14 F | ≥ 12 | 38 |
9-ACP2-010-034 | 34 | 28.0-31.0 | 10 | 14 F | ≥ 12 | 41 |
Contents: 1 each occluder, 145 cm delivery cable, plastic vise for delivery cable, hemostasis valve, loader, loading cable and plastic vise for loading cable, sheath adaptor (16 – 25 mm devices only).
MAT-2006710 v2.0
You are about to enter an Abbott country- or region-specific website.
Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions