Go Beyond the Angiogram

Comprehensive Physiology Indices

Go Beyond the Angiogram

Why Using Physiological Indices Is Essential

Physiological measurements, including hyperemic (FFR) and non-hyperemic pressure ratios (NHPR), Index of Microcirculatory Resistance (IMR), and Coronary Flow Reserve (CFR) are used to assess for epicardial disease and microvascular dysfunction. Physiologic data from PressureWire™ X Guidewire leads to informed decision-making that is proven to improve patient outcomes and prevent unnecessary, repeat hospitalizations.1

Physiological Indices Defined

Objective Diagnosis of Epicardial Disease and Microvascular Dysfunction

View how PressureWire™ X Guidewire measures Fractional Flow Reserve (FFR), Resting Full-cycle Ratio (RFR), Index of Microcirculatory Resistance (IMR) and Coronary Flow Reserve (CFR).

PressureWire X Guidewire
Assessing the Microcirculation

Test your knowledge of physiological indices with our quiz

 Physiology Indices

Physiology-Guided Decision Making Leads to Better Outcomes

Determining the Need for Revascularization
Determining the Need for Revascularization
Assessment for a Coronary Microvascular Dysfunction Diagnosis
Assessment for a Coronary Microvascular Dysfunction Diagnosis

Society Guidelines for Physiologic Indices

 Diagnosing Epicardial StenosisDiagnosing Coronary Microvascular Dysfunction
 FFRRFRIMRCFR
GuidelinesACC/SCAI (view here)
AHA/ACC Guidelines (view here)
Cutoff Value≤ 0.82≤ 0.893≥ 251,*< 2.51,*
FormulaPd/Pa at hyperemiaLowest Pd/Pa ratio of whole-cyclePd x Hyperemic FlowHyperemic Flow/Baseline Flow
HyperemiaYesNoYesYes

*IMR and CFR cutoffs in population of ischemia with no obstructive coronary artery disease (INOCA) patients, CorMicA trial.1

RFR is recognized by the Appropriate Use Criteria with an approved coding recommendation by ACC/SCAI.

Assessing whether Epicardial Stenosis needs intervention because there's more to CAD than an angiogram shows

Assessing Coronary Microvascular Dysfunction (CMD) because CMD may contribute to persistent angina

References

  1. Ford TJ, et al. 1-year outcomes of angina management guided by invasive coronary function testing (CorMicA). J Am Coll Card Intu. 2020;13:33-45.
  2. Jeremias A, et al. A test in context: fractional flow reserve: accuracy, prognostic implications, and limitations. J Am Coll Cardiol. 2017;69(22):2748-2758.
  3. Kobayashi Y, et al. Agreement of the resting distal to aortic coronary pressure with the instantaneous wave-free ratio. J Am Coll Cardiol. 2017;70(17):2105-2113.

MAT-2310906 v1.0

Important Safety Information
 

Coroventis CoroFlow Cardiovascular System

rx-only-logo

Indications: CoroFlow is indicated to provide hemodynamic information for use in the diagnosis of patients with cardiovascular diseases.

CoroFlow is intended for use in catheterization and related cardiovascular specialty laboratories to compute and display various physiological parameters based on the output from one or more measuring devices.

Contraindications: The system has no patient alarm functions. Do not use for cardiac/vital signs monitoring.

Warnings:

  • If CoroFlow is used together with 3rd party infusion catheters for assessment of Absolute Flow and Resistance, ensure that the maximum infusion rate per manufacturers instruction is not exceeded or vessel injury may occur.
  • Do not use the CoroFlow measurement system if there is reason to believe the system's security has been compromised or if the system was unaccounted for a period of time (i.e. misappropriated, modified or tampered with).
  • Do not leave the CoroFlow measurement system unattended when logged in as a PC Administrator
  • To protect the privacy and security of sensitive information, including electronic protected health information (EPHI), and to protect the integrity of the system itself, the system should be located in a physically secure, access-controlled environment.
  • To protect the privacy and security of sensitive information, including electronic protected health information (EPHI), the PC on to which CoroFlow is installed must be configured according to the Installation Instructions in this manual. Failure to configure the PC correctly may result in increased risk for unauthorized release of protected health information. Windows settings include:
    • Activation and configuration of restricted user Access
    • Activation of Windows Firewall and blocking of network connections
    • Activation of Windows Bitlocker drive encryption
    • Activation of Windows Secure Boot
    • Activation of Windows Anti-Virus scanning
    • Activation of Windows update
  • Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  • Use of accessories, transducers and cables other than those specified or provided by Coroventis could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  • Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of CoroFlow, including cables specified by Coroventis. Otherwise, degradation of the performance of this equipment could result.

 

Precautions:

  • The PC and CoroHub shall not be placed within the patient environment (1.5 m from patient).
  • For operation of other devices used in conjunction with CoroFlow consult the IFU for each of these devices for details on indication, handling and safety information.
  • It is recommended to ensure local routines for data backup of stored recordings. CoroFlow does not create backup of stored data.
  • Always check minimum performance requirement on PC to ensure compatibility with CoroFlow.
  • It is recommended to install CoroFlow on a PC with backup battery to avoid interruption in case of power failure.
  • Always manually review and confirm valid cursor positions and detected heart beats.
  • Ensure that Pa and Pd pressure waveforms are aligned in phase and offset after equalization, or indices can be mis-calculated.
  • Confirm that the correct Wi-Box is selected by manually matching the Wi-Box ID number with the Wi-Box in the lab.
  • Changing parameter settings outside of default values may affect measurement performance, only for research purposes.
  • Only to be used by healthcare professionals
  • Using a network location to store data may cause previously unidentified risks if the network malfunctions
  • The assembly of medical electrical systems and modifications during the actual service life require evaluation to the requirements according to IEC 60601-1 standard series.
  • CoroHub does not have any serviceable parts and require no field maintenance. No modification or tampering with CoroHub is permitted.
  • CoroHub shall not be immersed in liquid.
  • CoroHub shall not be used if it has been subject to damage
  • Direct connection to a non-secure network, like the internet, may interfere with correct operation and/or result in inappropriate access to patient information. Furthermore, it should be noted that reconfiguring a used network may lead to inability to import patient as well as export examination data, ultimately leading to a risk of loss of patient and examination data. To avoid this problem Coroventis recommends verifying network settings in the system setup after each change.

MAT-2007904 v3.0

PressureWire™ X Guidewire

rx-only-logo

Indications: The PressureWire™ X Guidewire is indicated to direct a catheter through a blood vessel and to measure physiological parameters in the heart and in the coronary and peripheral blood vessels. Physiological parameters include blood pressure. The PressureWire™ X Guidewire can also measure blood temperature.

Contraindications: This guidewire is contraindicated for use in the cerebral vasculature.

Warnings:

  • No modification of this device is allowed.
  • The PressureWire™ X Guidewire is supplied sterile. Discard the guidewire if the pouch is opened or damaged, compromising the sterile barrier. The guidewire is designed for single use only and shall not be reused or resterilized. Adverse effects of using a non-sterile or resterilized guidewire may include, but are not limited to:
    • Local and/or systemic infection
    • Mechanical damage
    • Inaccurate readings
  • Observe all guidewire movements. Whenever the guidewire is moved or torqued, the tip movement should be examined under fluoroscopy. Never push, withdraw, or torque the guidewire if it meets resistance or without observing corresponding movement of the tip, otherwise vessel/ventricle trauma may occur.
  • Torquing or excessive manipulation of the guidewire in a sharp bend, against resistance, or repeated attempts to cross a total vessel occlusion may:
    • Cause dissection or perforation of blood vessels
    • Cause vessel spasm
    • Damage and/or fracture the guidewire
  • When introducing the guidewire, flush the catheter and administer anticoagulation as for a standard catheterization procedure or clotting may occur.
  • Do not use the guidewire in the ventricles if the patient has a prosthetic mechanical or biological valve. It may result in damage to both the prosthesis and the guidewire, which may cause injury or death.
  • Use of the PressureWire™ X Guidewire in conjunction with interventional devices with a short rapid exchange may result in a folded or fractured guidewire.
  • High frequency surgical devices must not be used on a patient at the same time as the guidewire.


Precautions:

  • The PressureWire™ X Guidewire is a delicate instrument and should be handled carefully.
  • Make sure that the transmitter is kept dry to ensure accurate pressure and/or temperature readings. Inaccurate readings may necessitate device replacement.
  • Do not use the guidewire in conjunction with atherectomy catheters. It may damage the guidewire.
  • Do not withdraw or manipulate the guidewire in a sharp-edged object. It may result in abrasion of the guidewire coating.
  • Factors that may affect the accuracy of the diagnostic information include, but are not limited to:
    • Improper placement of the aortic pressure sensor.
    • Failure to achieve maximum coronary and myocardial hyperemia in FFR procedures.
    • Blood flow affected by the position of interventional devices, such as balloon catheters.
  • Guidewire readings may be affected by defibrillation. Rezero the guidewire after defibrillation use.
  • Do not measure pressure when the guidewire sensor element is in a sharp bend or in contact with atrial or ventricular walls. It might result in pressure artifacts.
  • Do not use the PressureWire™ X Guidewire together with another guidewire, for so called jailed wire technique, due to difficulty in guidewire withdrawal and possible guidewire entrapment.
  • Store at room temperature (15°C – 25°C) in a dry and dark place.


Potential Adverse Events: Potential complications which may be encountered during all catheterization procedures include, but are not limited to: vessel dissection or occlusion, perforation, embolus, spasm, local and/or systemic infection, pneumothorax, congestive heart failure, myocardial infarction, hypotension, chest pain, renal insufficiency, serious arrhythmias, or death.

In addition, this device has a coating containing Polyethylene Glycol (PEG); potential allergic reactions (anaphylaxis) may occur during the interventional procedure if the patient is allergic to PEG.