Product Advisories

Medical Device Correction: 2nd Generation CentriMag Primary Console Surge Immunity - May, 2024

Abbott is notifying users of a potential issue with the CentriMag™ Console (2nd Generation CentriMag Primary Console) where if subjected to a power surge above 1.8kV and up to 2.0 kV, the CentriMag Console may shut down completely without alarm. The letter contains important information and recommendations on how to use CentriMag Console if the system shuts down without an alarm.

Urgent Medical Device Recall: CentriMag Acute Circulatory Support System - August 22, 2019

An urgent medical device recall notice to customers communicating that we have received reports of CentriMag™ Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI).

Urgent Medical Device Recall: CentriMag Acute Circulatory Support System – September 5, 2018

An urgent medical device recall notice to customers communicating that we have received reports where damage to the cable that connects the motor to the console resulted in interruption of active support. This communication is to remind our physician and hospital partners of the importance of following the proper method for exchanging the CentriMag Acute Circulatory Support System Motor Model 102956 in the event the system experiences an interruption of power.