Armada™ 14 Balloon Dilatation Catheter

Ordering Information

Armada™ 14 PTA Catheter
Stock NumberDiameter (mm)Balloon Length (mm)Sheath CompatibilityInflation Pressure (atm)
90 (cm)150 (cm)Nominal/RBP
A1015-020A2015-0201.52048 / 14
A1015-040A2015-0401.54048 / 14
A1015-080A2015-0801.58048 / 14
A1015-120A2015-1201.512048 / 14
A1020-020A2020-0202.02048 / 14
A1020-040A2020-0402.04048 / 14
A1020-060A2020-0602.06048 / 14
A1020-080A2020-0802.08048 / 14
A1020-120A2020-1202.012048 / 14
A1020-200
A2020-2002.020048 / 14
A1025-020A2025-0202.52048 / 14
A1025-040A2025-0402.54048 / 14
A1025-060A2025-0602.56048 / 14
A1025-080A2025-0802.58048 / 14
A1025-120A2025-1202.512048 / 14
A1025-200A2025-2002.520048 / 14
A1030-020A2030-0203.02048 / 14
A1030-040A2030-0403.04048 / 14
A1030-060A2030-0603.06048 / 14
A1030-080A2030-0803.08048 / 14
A1030-120A2030-1203.012048 / 14
A1030-200A2030-2003.020048 / 14
A1040-020A2040-0204.02048 / 14
A1040-040A2040-0404.04048 / 14
A1040-060A2040-0604.06048 / 14
A1040-080A2040-0804.08048 / 14
A1040-120A2040-1204.012048 / 14
A1040-200A2040-2004.020048 / 14

 

Armada™ 14 XT PTA Catheter
Stock NumberCatheter Length (cm)Diameter (mm)Balloon Length (mm)Sheath Compatibility (F)Inflation Pressure (atm)
Nominal/RBP
F2012-0121451.21248 / 14
F2012-0201451.22048 / 14
F2015-0121451.51248 / 14
F2015-0201451.52048 / 14
F2020-0121452.01248 / 14
F2020-0201452.02048 / 14

 

Data on file at Abbott

MAT-2114519 v1.0

Important Safety Information

Armada™ 14
PTA Catheter

rx-only-logo

Indications

The device is indicated to dilate stenoses in femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The 2.0 to 4.0 mm balloon diameters are also indicated for post-dilatation of balloon-expandable stents up to 40 mm and self-expanding stents up to 80 mm in the vessels listed above.

Contraindications

  • Inability to cross lesion with a guide wire
  • Use in the coronary arteries

Warnings / Precautions

  • This device should only be used by physicians who are experienced and have a thorough understanding of the clinical and technical aspects of PTA.
  • One-time use only – do not resterilize! This single use device cannot be reused on another patient, as it is not designed to perform as intended after the first usage. Changes in mechanical, physical, and/or chemical characteristics introduced under conditions of repeated use, cleaning, and/or resterilization may compromise the integrity of the design and/or materials, leading to contamination due to narrow gaps and/or spaces and diminished safety and/or performance of the device. Absence of original labeling may lead to misuse and eliminate traceability. Absence of original packaging may lead to device damage, loss of sterility, and risk of injury to patient and/or user.
  • Do not use if inner package is damaged or opened.
  • Employ aseptic techniques during removal from the package and during use.
  • Any use for procedures other than those indicated in these instructions is not recommended.
  • Use prior to the use by date.
  • Carefully inspect the catheter prior to use to verify that it has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.
  • Precautions to prevent or reduce blood clotting should be taken when any catheter is used.
  • Flush or rinse all products entering the vascular system with sterile isotonic saline or a similar solution via the guide wire access port prior to use. Consider the use of systemic heparinization.
  • When the system is introduced into the vascular system, it should be manipulated only under high quality fluoroscopy.
  • The Armada™ 14 PTA Catheter must always be introduced, moved and or withdrawn over a guide wire (max. 0.014″).
  • Never attempt to move the guide wire when the balloon is inflated.
  • Never use air or any gaseous medium to inflate the balloon.
  • Do not advance the Armada™ 14 PTA Catheter against significant resistance. The cause of resistance should be determined via fluoroscopy and remedial action taken.
  • The minimal acceptable sheath French size is printed on the package label. Do not attempt to pass the Armada™ 14 PTA Catheter through a smaller sized sheath introducer than indicated on the label.
  • The size of the inflated balloon should be selected not to exceed the diameter of the artery immediately distal, or proximal, to the stenosis.
  • Inflation in excess of the rated burst pressure may cause the balloon to rupture. Use of a pressure monitoring device is recommended.
  • If a distal protection device is used, follow the manufacturer’s instruction for use. Allow and maintain adequate distance between the Armada™ 14 PTA Catheter and the distal protection device to avoid engagement.
  • Rated burst pressure and balloon fatigue testing of the Armada™ 14 PTA balloons within deployed stents has demonstrated the following:
    • The 2.0 to 4.0 mm balloon diameters can safely post-dilate balloon expandable stents up to 40 mm in length.
    • The 2.0 to 4.0 mm balloon diameters can safely post-dilate self-expanding stents up to 80 mm in length.

The safety of using additional balloon diameters and/or lengths to post dilate stents has not been established.

  • When post-dilating stents, use a balloon length that is appropriate for the deployed stent length.

Potential Complications

The following complications may occur as a result of PTA, but may not be limited to:

  • Abrupt closure
  • Access site hematoma
  • Aneurysm
  • Angina
  • Arrhythmias
  • Arteriovenous fistula
  • Bleeding complications which may require transfusion
  • Cerebral ischemia/transient ischemic attack (TIA)
  • Death
  • Embolism (air, tissue, thrombotic, systemic or device component)
  • Fever/pyrogenic reaction
  • Hypersensitivity or allergic reaction to contrast agents and drug reactions
  • Hypertension / hypotension
  • Infection
  • Ischemia, including tissue ischemia, steal syndrome and necrosis
  • Leg edema
  • Myocardial ischemia or infarction
  • Nausea and vomiting
  • Neuropathies or nerve injury
  • Occlusion
  • Organ failure (single, multiple)
  • Pain
  • Palpitations
  • Pseudoaneurysm
  • Renal failure/insufficiency
  • Restenosis
  • Stroke / cerebrovascular accident (CVA)
  • Vascular complications, including entry site, which may require vessel repair
  • Vascular thrombosis
  • Vessel injury, e.g. dissection, perforation
  • Vessel spasm

 

 

MAT-2114592 v2.0