Experience indicates that certain models of implantable cardiac defibrillators (ICDs) and certain implantable pacemakers (IPMs) may, in some cases, not be able to establish telemetry and be reprogrammed due to electromagnetic interference when used with the HeartMate 3™ LVAS.
In such cases the ICDs or IPMs have continued to function properly and only their ability to communicate with the programmer was affected. The HeartMate 3 LVAS complies with applicable electromagnetic compatibility standards and is not influenced by such devices. No such difficulties have been reported with models other than those listed below.
Difficulty or inability to communicate with the ICD or IPM programmer has been reported for or may occur with the following manufacturer families:
Manufacturer | Family / Model |
---|---|
Biotronik | Acticor/Rivacor |
Biotronik | Ilivia Neo/Intica Neo |
Biotronik | Ilivia/Intica/Inlexa |
Biotronik | Itrevia/Iperia/Inventra |
Biotronik | Iforia/Ilesto/Idova |
Biotronik | Lumax |
ELA Medical (Sorin) | Paradyme CRT |
Medtronic | Cardia CRT |
Boston Scientific | ORIGEN CRT |
MAT-2105186 v2.0
You are about to enter an Abbott country- or region-specific website.
Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions