Low-profile embolisation system compatible with 0.038" guidewire-compatible diagnostic catheters¹ designed to access distal vasculature through tortuous anatomy.²
A - Vascular Plug Diameter
B - Unconstrained Length
Amplatzer™ Vascular Plug 4
Vessel Size | Device Dimensions | Diagnostic Catheter Requirements* | |||
---|---|---|---|---|---|
Treatable Vessel Diameter Range† | Model/Reorder Number | Vascular Plug Diameter (A) | Unconstrained Length (B) | Requirements for diagnostic catheters** | Maximum Diagnostic Catheter Length*** |
2.5 mm-3.0 mm | 9-AVP038-004 | 4 mm | 10.0 mm | 0.038" Guidewire-Compatible Diagnostic Catheter | ≤ 125 cm |
3.5 mm-4.0 mm | 9-AVP038-005 | 5 mm | 10.5 mm | 0.038" Guidewire-Compatible Diagnostic Catheter | ≤ 125 cm |
4.0 mm-4.5 mm | 9-AVP038-006 | 6 mm | 11.0 mm | 0.038" Guidewire-Compatible Diagnostic Catheter | ≤ 125 cm |
4.5 mm-5.5 mm | 9-AVP038-007 | 7 mm | 12.5 mm | 0.038" Guidewire-Compatible Diagnostic Catheter | ≤ 125 cm |
5.5 mm-6.0 mm | 9-AVP038-008 | 8 mm | 13.5 mm | 0.038" Guidewire-Compatible Diagnostic Catheter | ≤ 125 cm |
†Treatable vessel diameter range based on the device's Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
*The Amplatzer™ Vascular Plug 4 is delivered using a Diagnostic Catheter meeting the minimum internal diameter requirements above.
**The Amplatzer™ Vascular Plug 4 has been tested for compatibility with the following diagnostic catheters and corresponding lengths:
5F Dx Catheters: Boston Scientific IMAGER‡ II (100 cm or less) and Merit Medical Impress‡ (125 cm or less)
4F Dx Catheters: Cordis‡ TEMPO‡ or TEMPO‡ AQUA‡ (100 cm or less)
***Each Amplatzer™ Vascular Plug 4 comes pre-loaded and packaged on an 0.038" x 155cm-long PTFE-coated delivery wire.
MAT-2413428 V1.0
Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions