Rapid embolisation2 and can adjust to variable landing zones3
A - Vascular Plug Diameter
B - Unconstrained Length
Amplatzer™ Vascular Plug II
Vessel Size | Device Dimensions | Delivery Catheter Requirements* | ||||||
---|---|---|---|---|---|---|---|---|
Treatable Vessel Diameter Range† | Model/Reorder Number | Vascular Plug Diameter (A) | Unconstrained Length (B) | Minimum and Maximum Internal Diameter | Minimum Sheath Size | or | Minimum Guide Catheter Size | Maximum Delivery Catheter Length** |
2.0 mm-2.5 mm | 9-AVP2-003 | 3mm | 6mm | 1.42-1.70 mm / 0.056-0.067 in | ≥4 F | ≥5 F | ≤ 100 cm | |
2.5 mm-3.0 mm | 9-AVP2-004 | 4mm | 6mm | 1.42-2.50 mm / 0.056-0.098 in | ≥4 F | ≥5 F | ≤ 100 cm | |
4.0 mm-4.5 mm | 9-AVP2-006 | 6mm | 6mm | 1.42-2.50 mm / 0.056-0.098 in | ≥4 F | ≥5 F | ≤ 100 cm | |
5.5 mm-6.0 mm | 9-AVP2-008 | 8mm | 7mm | 1.42-2.69 mm / 0.056-0.106 in | ≥4 F | ≥5 F | ≤ 100 cm | |
6.5 mm-7.5 mm | 9-AVP2-010 | 10mm | 7mm | 1.78-2.69 mm / 0.070-0.106 in | ≥5 F | ≥6 F | ≤ 100 cm | |
8.0 mm-9.0 mm | 9-AVP2-012 | 12mm | 9mm | 1.78-2.69 mm / 0.070-0.106 in | ≥5 F | ≥6 F | ≤ 100 cm | |
9.5 mm-11.0 mm | 9-AVP2-014 | 14mm | 10mm | 2.18-2.69 mm / 0.086-0.106 in | ≥6 F | ≥8 F | ≤ 100 cm | |
10.5 mm-12.5 mm | 9-AVP2-016 | 16mm | 12mm | 2.18-2.69 mm / 0.086-0.106 in | ≥6 F | ≥8 F | ≤ 100 cm | |
12.0 mm-14.0 mm | 9-AVP2-018 | 18mm | 14mm | 2.49-2.69 mm / 0.098-0.106 in | ≥7 F | ≥9 F | ≤ 100 cm | |
13.5 mm-15.5 mm | 9-AVP2-020 | 20mm | 16mm | 2.49-2.69 mm / 0.098-0.106 in | ≥7 F | ≥9 F | ≤ 100 cm | |
14.5 mm-17.0 mm | 9-AVP2-022 | 22mm | 18mm | 2.49-2.69 mm / 0.098-0.106 in | ≥7 F | ≥9 F | ≤ 100 cm |
†Treatable vessel diameter range based on the devices Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
*The Amplatzer™ Vascular Plug II is delivered utilizing either a sheath or guide catheter meeting the minimum internal diameter requirements above.
**Each Amplatzer™ Vascular Plug II comes pre-loaded on a 135cm nitinol delivery wire.
MAT-2413425 V1.0
Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions